Current Probe Calibration Generators KSZ
■ KSZ 100D
■ 20A to 100A
■ 2% Accuracy
■ Rise time 15ns (20A) - 80ns (100A)
■ Manual or Remote Operation
■ Optional Interlock
The KSZ calibration generator series is used for RF and DC control of current clamps and current transformers up to 100A. Suitable current probes are for example our IWATSU® Rogowski type probes, and our IWATSU® AC/DC current clamps.
The KSZ100D includes remote control capabilities.
To use the current calibration generator, a current clamp must be clamped to the current bus bar. The current bus bar is configured such that it is removeable. This allows connecting closed-loop current transformers, like our LILCO® series.
Our current transformers can be contacted by a current line at the safety sockets.
Before delivery, the KSZ calibration generators are compared in terms of pulse and rise behavior with calibrated instruments whose calibration is traceable to DKD standards.
Each calibration generator is delivered factory calibrated.
Order number | Article | Output | Pulse Current Accuracy | Rise time | |||
---|---|---|---|---|---|---|---|
894-010-D00 | KSZ 100 D | 20A / 50A / 100A | 2% | 15 - 80 ns |
Options
INTERL-KSZ Interlock option for all KSZ models with control cable (2m) for connection to a closing contact (protective cover). Connection to the device via LEMO-Push-Pull connector on the back of the device.
Order number | Article | |||
---|---|---|---|---|
|
INTERL-KSZ | KSZ Option: Interlock (for KSZ 10 and KSZ 100 series) |
High-Current Calibration Generators
1 product